Technical Data
12.2 Defibrillation Waveform
Page 136
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
Safety discharge when:
• the battery voltage is insufficient
• the shock is not released within 20 seconds
• a lower energy value is selected while the defibrillator is charging
• the device is defective
• the device is turned off
• a non shockable rhythm is detected
Shock delivery
• Via applied disposable adhesive defibrillation electrodes
Defibrillation electrode
connection
Type BF, defibrillation-protected >5 kV
Defibrillation electrodes
Adult and Paediatric electrode
Electrode cable, 2 m long
Deviating from the compliance statement according to 201.108.1.10 (IEC60601-2-
4, 201.1.108.7. and 201.1.108.6), the following features have been measured for the
universal electrode:
• 80 cm2 active surface
Measurements after 60 minutes of pacemaker at maximum setting, followed by a
shock at 200 Joule:
• Defibrillation recovery 720 mV after 4 s and 710 mV after 60s
• DC offset voltage 785 mV after 1 minute of stabilisation