Technical Data
12.4 Technical data - monitoring
Page 144
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12.4.7
etCO
2
- Capnography
Extremely compact infrared mainstream and sidestream gas analyser.
The IRMA and ISA sensors are equipped with an automatic barometric pressure
compensation.
Trademarks
Masimo IRMA
TM
Masimo ISA
TM
,
Nomoline
TM
Masimo AB patents
SE519766; SE519779; SE523461; SE524086
Standards
MDD 93/42/EEC, EN ISO 80601-2-55:2011, IEC 60601-1:2005, IEC 60601-1-
2:2007, EN ISO 5356-1:2004, EN 1789:2007
Modules
Masimo IRMA mainstream and Masimo ISA sidestream
Cable length
2.5 m (IRMA)
0.5 m (ISA)
Surface temperature IRMA
(at environment temp. of 23° C)
Max. 39 °C
Accuracy
in standard conditions
in all conditions
The following specifications are valid for a dry gas at 22 ± 5°C and 1013 ± 40 hPa
(standard conditions).
• 0...15 % (± 0.2 vol % + 2 % of reading)
• 15... 25 % (no information on accuracy)
• ± (0.3 kPa + 4 % of reading)
Specification of accuracy in all conditions is valid for all specified ambient conditions (see
page
). Deviating from these, influences are specified in the table “Interfering gas and
vapour effects” and in section “Effects from water vapor partial pressure on gas readings”.
These technical specification can be found in the IRMA/ISA user guide chapter 2.
Breath detection
Adaptive threshold value, min. 1 vol% change in CO2 concentration
Respiration rate
0-150 /min. The respiration rate is displayed after three breaths and the average
value is updated after every breath.
Rise time
IRMA: CO
2
< 90 ms
ISA: CO
2
< 500 ms