4
Monitoring
4.8
IBP Monitoring
Page 60
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
4.8.2
Start IPB measurements
IBP curve display
4.8.3
IBP menu settings
Access the IBP menu via the IBP measurement field as described on page
.
The default settings are printed
bold
.
1. Select the IBP measurement
field
(1)
to open the IBP menu.
2. Select OFF/ON button
(2)
to start the measurement.
3. Zeroing the IBP (see Zeroing)
4. Check the IBP curve on the display to see if the connections have been made
correctly and the IBP value is in the expected range.
Start measurement
ON
1
Curve amplitude (mmHg)
Zeroing
2
IBP curve
IBP measurement field with
Systolic/Diastolic and mean
arterial pressure
Screenshot
Manual Def
Start
Menu
Event
R-ECG
HR
Menu item
Parameter
Description
IBP parameter
Start measurement
Starts the IBP measurement
Zeroing
Zero adjustment of the IBP
Limits
see Alarm limits
Curve amplitude (mmHg)
Set the range for the IBP measurement:
0-30, 0-60, 0-150, 0-300 mmHg