Technical Data
12
User guide
Device Configuration
12.6
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12.6.3
Defibrillator
12.6.4
Display
Parameter
Values
Description
Default energy for
adults in manual def.
• 2, 4, 8, 15, 30, 50, 70, 90,
120,
150*
, 200 Joule
Sets the energy which will be displayed by default when entering
manual defibrillation mode in adult mode
Default energy for
children in manual def.
• 2, 4, 8, 15, 30,
50*
, 70, 90,
Joule
Sets the energy which will be displayed by default when entering
manual defibrillation mode in child mode
Sync. after sync shock
• True
• False*
If this option is activated, the device remains in sync. mode after
a synchronized shock in manual defibrillation
First shock for adults
• 2, 4, 8, 15, 30, 50, 70, 90,
120,
150
, 200 Joule
Sets the energy which will be delivered for the first shock in AED
in adult mode
Second shock for adults
• 2, 4, 8, 15, 30, 50, 70, 90,
120,
150
, 200 Joule
Sets the energy which will be delivered for the second shock in
AED in adult mode
Third shock for adults
• 2, 4, 8, 15, 30, 50, 70, 90,
120,
150
, 200 Joule
Sets the energy which will be delivered for the third shock in AED
in adult mode
First shock for children
• 2, 4, 8, 15, 30,
50*
, 70, 90,
Joule
Sets the energy which will be delivered for the first shock in AED
in child mode
Second shock for
children
• 2, 4, 8, 15, 30,
50*
, 70, 90,
Joule
Sets the energy which will be delivered for the second shock in
AED in child mode
Third shock for children
• 2, 4, 8, 15, 30,
50*
, 70, 90,
Joule
Sets the energy which will be delivered for the third shock in AED
in child mode
Parameter
Values
Description
Anteriority Analysis
• False*
• True
Enables the analysis with anteriotity (part of the signal analysis is done
on the signal before the device mentions the start of the analysis).
Analysis key
• False
•
True*
If this option is activated, the analysis key will be displayed to allow
additional manual analysis in AED mode.
Alarm on Vfib or
Vtach detection
• False*
•
True*
High priority alarm will be issued if Vfib or Vtach is detected during
monitoring on both ECG cable and Defi pads.
Precognitive
Analysis
• False*
• True
Not used at the moment, has no impact on the unit.