The dwell time refers to the time of dwelling at the frequency point exceeding the set limit in the ON state.
The schematic diagram of FScan mode is shown in the figure below (example).
Fig. 9-2 Schematic Diagram of FScan Mode
3
MScan
Main operation steps of MScan:
1) Press [Edit List]→ [Edit List] and edit the current list.
2) Press [Measure]→[MScan] to enable the MScan mode.
3) Press [Frequency]→ [Step Freq] and set the step frequency of sweeping.
4) Press [Frequency]→ [Points] and set the number of sweeping points.
5) Press [BW]→ [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz,
50kHz, 120kHz or 150kHz.