8
Audio spectrum menu
Span
Audio Spectrum
Scale/Div
10.0dB
Ref Position
· [Span]:
Press [Audio Spectrum]→ [Span], and set the span of the
audio spectrum. The span should be large enough to display the
audio signal and related harmonics. You can change the span with
the number keys and select the frequency unit, or with the [↑] or [↓]
key or knob.
· [Scale/Division]:
Press [Audio Spectrum]→ [Scale/Div], and set
the scale of the audio spectrum, so as to facilitate the observation of
the audio spectrum trace. You can change the scale unit with the
number keys and select the power unit [dB] or [-
dB], or use the [↑] or
[↓] key or knob.
· [Reference position]:
Press [Audio Spectrum]→ [Ref Position],
and set the reference position of the audio spectrum to facilitate the
observation of the audio spectrum line. You can change the reference
position with the number keys and click [OK], or with the [↑] or [↓]
key or knob.
9
Audio waveform menu
Sweep Time
Audio Waveform
Scale/Div
20.0%
· [Sweep time]:
Press [Audio Waveform]→ [Sweep Time] and set
the audio waveform sweeping time. The longer the sweep time is and
the more points are sampled, the stabler the measurement results will
be.
· [Scale/Division]:
Press [Audio Waveform]→ [Scale/Div] and set
the scale of the audio waveform to facilitate the observation of the
audio spectrum trace. You can change the scale unit with the number
keys and select [%] or click [OK], or with the [↑] or [↓] key or knob.
· [Special note]:
The scale unit changes with the type of the
modulation signal. The scale unit should be percentage (%) in
AM signal measurement, frequency unit (Hz, kHz or GHz) in
FM signal measurement and radian (Rad) in PM signal
measurement.
Содержание S3302A
Страница 1: ...S3302 Series Handheld Spectrum Analyzer Datasheet Saluki Technology Inc...
Страница 70: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 70...
Страница 125: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 125 Fig 7 1 FM Demodulation Analysis Results...
Страница 139: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 139 Fig 8 3 Schematic Diagram of List Scanner...