· [Auto]:
Press [BW]→ [Detector]→ [Auto] to enable the spectrum analyzer
mode of the detector.
· [Normal]: Press [BW]→ [Detector]→ [Normal] to enable the most
commonly used detector mode. You can see the signal and noise floor at
the same time, with no signal loss.
· [Peak]: Press [BW]→ [Detector]→ [Peak] to prevent missing of any peak
signal. This function can be applied to measure the signal close to the
noise floor.
· [Neg Peak]: Press [BW]→[Detector]→[Neg Peak]. This function is used
in self-inspection of the MMW integrated test instrument in most cases and
rarely in tests. By using this function, the modulation envelope of AM signal
can be well reproduced.
· [Sample]: Press [BW]→ [Detector]→ [Sample]. This function is
conducive for noise signal measurement. Compared with the normal mode
of the detector, noise can be better measured.
· [Average]: Press [BW]→ [Detector]→ [Average] to average the data
within the sampling interval.
· [RMS]: Press [BW]→ [Detector]→ [RMS] to obtain the RMS value of the
data within the sampling interval.