Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
55
Press [Amplitude] and [Ref Level], and set it as the peak power. Results are shown in Fig. 4-18.
Fig. 4-18 Input Signal under 100kHz Span
g) Set the step of the center frequency as the signal frequency of the fundamental wave.
Press [Frequency] and [Frequency Step Auto ON/OFF] and enter 1 GHz.
h) Measure the second harmonic.
Press [Maker], [Marker→], [Marker→Center] and step key [↑]. Change the center frequency of the spectrum
analyzer to second harmonic by stepping operation. Press [Peak] and [Peak Search], and read the peak power.
Press [Amplitude] and [Ref Level], and set it as the peak power. Adjust the harmonic peak to the reference level.
The second harmonic amplitude is shown in Fig. 4-19.