IQ Capture
Off On
Start Capture
IQ Capture
Capture Time
1.000ms
Capture Mode
Single
Continuous
Sample Rate
5.000MHz
Triggering
[Free Run]
Save Name
[IQCapt...]
< Back
>
· [IQ Capture Off On]: Press [Measure]→ [More 1/2]→ [IQ Capture]→
[IQ Capture Off On] to enable or disable the IQ capture function.
· [Start Capture]: Press [Measure]→ [More 1/2]→ [IQ Capture]→ [Start
Capture] to start IQ capture.
· [Capture time]: Press [Measure]→ [More 1/2]→ [IQ Capture]→
[Capture Time] to set the IQ capture time.
· [Capture Mode]: Press [Measure]→ [More 1/2]→ [IQ Capture]→
[Capture Mode] to enable the single or continuous IQ capture. In the single
mode, the data will be captured once only. In the continuous mode, the
data will be captured once after each sweeping of the stroke, and capturing
will not be stopped until otherwise set by the user.
· [Sample rate]: Press [Measure]→ [More 1/2]→ [IQ Capture]→ [Sample
Rate] to set the sample rate of IQ capture.
· [Triggering]: Press [Measure]→ [More 1/2]→ [IQ Capture]→
[Triggering] to set the triggering mode, including [Free Run] and
[External]. In the [External] mode, [Slope] and [Delay] can be set.
· [Save Name]: Press [Measure]→ [More 1/2]→ [IQ Capture]→ [Save
Name] to save the captured data.
Содержание S3302A
Страница 1: ...S3302 Series Handheld Spectrum Analyzer Datasheet Saluki Technology Inc...
Страница 70: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 70...
Страница 125: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 125 Fig 7 1 FM Demodulation Analysis Results...
Страница 139: ...Tel 886 2 2175 2930 Email sales salukitec com www salukitec com 139 Fig 8 3 Schematic Diagram of List Scanner...