Remote Control Commands
R&S
®
NRPxxTWG
49
User Manual 1178.8371.02 ─ 04
TRIGger:EXTernal<2...2>:IMPedance
................................................................................ 51
TRIGger:ATRigger:DELay
<delay>
Effective only if
ON
.
Sets the delay between the artificial trigger event and the beginning of the actual mea-
surement
Parameters:
<delay>
Range:
0.1 to 5.0
*RST:
2.0
Default unit: Seconds
TRIGger:ATRigger:EXECuted?
Queries the number of measurements that were triggered automatically since
was set to
ON
.
In normal scalar measurements, this number can only be 0 or 1. If a buffered measure-
ment was executed, this number indicates how many results in the returned array of
measurement data were executed without a real trigger event.
Usage:
Query only
TRIGger:ATRigger:STATe
<state>
Controls the automatic trigger function. If enabled, an artificial trigger is generated if the
delay time has elapsed after the measurement start and no trigger event has occurred.
The delay time is set using
Parameters:
<state>
*RST:
OFF
Controlling the Measurement