Remote Control Commands
R&S
®
NRPxxTWG
90
User Manual 1178.8371.02 ─ 04
bb
= blue
Range:
0x00 to 0x02FFFFFF
*RST:
0x00A0A0A0
Example:
SYST:LED:MODE USER
Sets the system status LED operating mode to user.
SYST:LED:COL #H01a00000
The LED flashes slowly in red.
SYSTem:LED:MODE SENSor
Sets the system status LED operating mode back to the sensor
internal settings.
SYSTem:LED:MODE
<mode>
Sets whether the color of the system status LED is controlled by the sensor firmware or
by the user settings.
For more information, see
.
Parameters:
<mode>
USER | SENSor
*RST:
SENSor
SYSTem:MINPower?
Queries the lower power measurement limit.
[SENSe<Sensor>:]CORRection:SPDevice:STATe
is set to
ON
. The lower measurement limit refers to the sensor or to the combination of a sensor
and the components connected ahead of it.
Use this query to determine a useful resolution for the result display near the lower
measurement limit.
Usage:
Query only
SYSTem:PARameters?
Lists all commands with default values, limits and ranges.
Usage:
Query only
SYSTem:PARameters:DELTa?
Lists all commands that differ from the defined default status set by
The commands are output with default values, limits and ranges.
Usage:
Query only
Configuring the System