Remote Control Commands
R&S
®
NRPxxTWG
61
User Manual 1178.8371.02 ─ 04
[SENSe<Sensor>:]AVERage:COUNt:AUTO
<state>
Sets the mode for determining the average count.
Parameters:
<state>
ON
Auto averaging: the averaging factor is continuously determined
and set depending on the level of power and other parameters.
OFF
Fixed filter: the previous, automatically determined averaging
factor is used.
ONCE
An averaging factor is determined by the filter automatic function
under the current measurement conditions and is then used in
the fixed filter mode.
*RST:
ON
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
<maximum_time>
Sets an upper limit for the settling time of the auto-averaging filter if
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
is set to
NSRatio
. Thus it limits
the length of the filter.
Parameters:
<maximum_time>
Range:
0.01 to 999.99
*RST:
4.00
Default unit: Seconds
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
<nsr>
Determines the relative noise component in the measurement result for the measure-
ment modes with scalar results. These measurement modes are Continuous Average,
Burst Average and Timeslot Average, provided the particular sensor supports them.
This command is only effective if the auto average calculation is enabled:
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
NSR
The noise component is defined as the magnitude of the level variation in dB caused
by the inherent noise of the sensor (two standard deviations).
The query returns the relative noise component in the measured value.
Parameters:
<nsr>
Range:
100.000e-6 to 1.00
*RST:
0.01
Default unit: dB
Configuring Basic Measurement Parameters