Operating Concepts
R&S
®
NRPxxTWG
28
User Manual 1178.8371.02 ─ 04
The measurement starts, and the result is displayed in dBm.
5. If necessary, perform further settings.
For a detailed description of how to measure in this setup, refer to the user manual of
the R&S NRX.
5.7
R&S NRP2
With the R&S NRPxxTWG power sensors and an R&S NRP2, you can measure power
with up to four power sensors simultaneously. All sensor-dependent measurement
functions can be used and the results can be displayed in parallel.
Required equipment
●
R&S NRPxxTWG power sensor
●
R&S NRP
‑
ZK6 cable to connect the sensor to the R&S NRP2
●
R&S NRP2 base unit with FW version 7.11 or higher
Setup
Figure 5-4: Setup with an R&S NRP2 base unit
1 = Signal source
2 = R&S NRPxxTWG power sensor
3 = Host interface connector
4 = R&S NRP
‑
ZK6 cable
5 = Sensor input connector of the R&S NRP2
6 = R&S NRP2 base unit
Incorrectly connecting/disconnecting the R&S NRPxxTWG power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Starting a measurement
1. Connect the cables as shown in
a) Connect the R&S NRP
‑
ZK6 cable to the host interface connector of the sensor.
R&S NRP2