Remote Control Commands
R&S
®
NRPxxTWG
67
User Manual 1178.8371.02 ─ 04
[SENSe<Sensor>:]CORRection:SPDevice:SELect
<num>
Selects a loaded data set for S-parameter correction.
See also
Chapter 7.8.3.3, "S-Parameter Correction"
Parameters:
<num>
Range:
1 to 1999
*RST:
1
Manual operation:
See
[SENSe<Sensor>:]CORRection:SPDevice:STATe
<state>
Activates or deactivates the S-parameter correction. If activated, uses the S-parameter
data set selected by
[SENSe<Sensor>:]CORRection:SPDevice:SELect
See also
Chapter 7.8.3.3, "S-Parameter Correction"
Parameters:
<state>
*RST:
OFF
Example:
CORR:SPD:SEL 1
Selects a S-parameter correction data set.
CORR:SPD:STAT ON
Activates the S-parameter correction.
Manual operation:
See
"S-Parameter Correction State Locked"
7.8.3.4
S-Gamma Corrections
Using the complex reflection coefficient, you can determine the power P delivered by
the signal source with considerably greater accuracy.
The coefficient of the signal source
Γ
source
is defined by its magnitude and phase:
●
[SENSe<Sensor>:]SGAMma:MAGNitude
●
The complex reflection coefficient
Γ
sensor
of the sensor, which is also required for the
correction, is prestored in the calibration data memory for many frequencies.
Sensor
Source
Figure 7-3: Correction of interactions between the power sensor and the signal source
Configuring Basic Measurement Parameters