Remote Control Basics
R&S
®
NRPxxTWG
119
User Manual 1178.8371.02 ─ 04
9.3.8.4
Operation Sense Status Register
The
CONDition
register contains information about whether a sensor is being initial-
ized and, depending on the configuration of the transition register. The
EVENt
register
indicates whether a sensor initialization was started or completed since the last read-
out of this register.
This status is assumed by a sensor if one of the following conditions is met:
●
The supply voltage is switched on (power up)
●
The sensor was connected
●
A reset was performed using:
–
–
Querying the register:
●
STATus:OPERation:SENSe:CONDition?
●
STATus:OPERation:SENSe[:SUMMary][:EVENt]?
Table 9-14: Meaning of bits used in the operation sense status register
Bit No.
Meaning
0
Not used
1
Sensor initializing
The sensor is being initialized.
5 to 15
Not used
9.3.8.5
Operation Lower Limit Fail Status Register
The
CONDition
register contains information about whether a displayed value is
below a configured lower limit. The
EVENt
register indicates whether a measured
value dropped below a limit value since the last readout of the Operation Lower Limit
Fail Status Register. Details of the behavior are defined by the transition register.
Querying the register:
●
STATus:OPERation:LLFail:CONDition?
●
STATus:OPERation:LLFail[:SUMMary][:EVENt]?
Table 9-15: Meaning of bits used in the operation lower limit fail status register
Bit No.
Meaning
0
Not used
1
Lower limit fail
The measured value drops below the lower limit value.
5 to 15
Not used