Remote Control Commands
R&S
®
NRPxxTWG
56
User Manual 1178.8371.02 ─ 04
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxTWG sets the last used
length.
Example for
REAL,32
format:
#6768000....<binary float values>....
Example for
REAL,64
format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for
ASCii,4
format:
1.2938e-06, -4.7269e-11, ...
*RST:
ASCii,0
7.6.3
Retrieving Results
After performing the measurement, you can query the measurement results with a
command from the
FETCh
subsystem.
Remote commands:
FETCh<Sensor>:ARRay[:POWer][:AVG]?
.......................................................................... 56
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
........................................................................56
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a measurement with enabled data buffer
mode.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result.
Usage:
Query only
7.7
Configuring the Measurement Modes
In the following chapter the settings needed for configuring a measurement mode are
described.
Configuring the Measurement Modes