Remote Control Commands
R&S
®
NRPxxTWG
59
User Manual 1178.8371.02 ─ 04
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast the
FETCh?
command returns a result only if the buffer is full.
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
BUFF:SIZE 1
See
Chapter 8.2, "Performing a Buffered Continuous Average
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer with
.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
BUFF:STAT OFF
[SENSe<Sensor>:][POWer:][AVG:]FAST
<state>
Enables or disables a fast unchopped Continuous Average measurement. The aver-
age count is enforced to 1. Any setting for average count is silently ignored. See also
"Calculating the measurement time"
Parameters:
<state>
*RST:
OFF
Example:
See
Chapter 8, "Performing Measurement Tasks - Programming
Configuring the Measurement Modes