Operating Concepts
R&S
®
NRPxxTWG
27
User Manual 1178.8371.02 ─ 04
Required equipment
●
R&S NRPxxTWG power sensor
●
R&S NRP
‑
ZK8 to connect the sensor to the R&S NRX
●
R&S NRX
Setup
Figure 5-3: Setup with an R&S NRX base unit
1 = Signal source
2 = R&S NRPxxTWG power sensor
3 = Host interface connector
4 = R&S NRP
‑
ZK8
5 = Sensor input connector of the R&S NRX
6 = R&S NRX base unit
Incorrectly connecting/disconnecting the R&S NRPxxTWG power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Starting a measurement
1. Preset the R&S NRX and the connected R&S power sensors.
a) Press the [Preset] key.
b) Tap "Preset".
All parameters are set to their defaults.
2. If measuring in zero-IF mode (RBW
> 40
MHz), consider to zero the power sensor:
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key of the R&S NRX.
c) Tap "Zero All Sensors".
3. Configure the measurement.
a) In the "Measurement Settings" dialog, select the "Measurement Type", for
example "Continuous Average".
b) Tap "Quick Setup"
> "Auto Set".
4. Switch on the signal source.
R&S NRX