Remote Control Commands
R&S
®
NRPxxTWG
81
User Manual 1178.8371.02 ─ 04
Example:
*CLS
CAL1:ZERO:AUTO ONCE
Performs zeroing. Takes several seconds.
SYST:SERR?
Query
-240
Response: Warning; Zero Calibration failed; Results Degrading.
7.10
Testing the Power Sensor
The selftest allows a test of the internal circuitry of the sensor.
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps
Offset Voltages
and/or
Noise Voltages
.
TEST:SENSor?
[<Item>]
Triggers a selftest of the sensor. In contrast to
, this command returns detailed
information that you can use for troubleshooting.
Query parameters:
<Item>
String
Usage:
Query only
7.11
Configuring the System
The
SYSTem
subsystem contains a series of commands for general functions that do
not directly affect the measurement.
The
SYSTem:COMMunicate:NETWork:...
commands are only available for the
R&S
NRP LAN power sensors.
Remote commands:
SYSTem:COMMunicate:NETWork:RESTart
........................................................................ 82
SYSTem:COMMunicate:NETWork:RESet
...........................................................................82
SYSTem:COMMunicate:NETWork:STATus?
....................................................................... 83
SYSTem:COMMunicate:NETWork[:COMMon]:DOMain
........................................................83
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
...................................................83
SYSTem:COMMunicate:NETWork:IPADdress
..................................................................... 84
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
.......................................................84
SYSTem:COMMunicate:NETWork:IPADdress:INFO?
...........................................................84
SYSTem:COMMunicate:NETWork:IPADdress:MODE
.......................................................... 84
Configuring the System