Remote Control Commands
R&S
®
NRPxxTWG
41
User Manual 1178.8371.02 ─ 04
Usage:
Query only
*WAI
WAIt to continue
Prevents the execution of the subsequent commands until all preceding commands
have been executed and all signals have settled.
Usage:
Event
7.4
Preparing for the Measurement
Before starting a measurement, you need to do the following:
Selecting the Reference Source
............................................................................. 41
............................................................................. 41
7.4.1
Selecting the Reference Source
The
ROSCillator
subsystem contains commands for configuring the reference
source.
[SENSe<Sensor>:]ROSCillator:SOURce
<source>
Sets the source of the reference oscillator. Refers typically to a precision, stabilized
time base.
Parameters:
<source>
INTernal | EXTernal | HOST
INTernal
Internal precision oscillator
EXTernal | HOST
External signal supplied at the host interface connector.
*RST:
INTernal
Example:
ROSC:SOUR INT
7.4.2
Selecting a Measurement Mode
► Before starting a measurement, select the measurement mode using:
The following modes are available:
●
Continuous average (
"POWer:AVG"
): After a trigger event, the power is integrated
over a time interval.
Preparing for the Measurement