Operating Concepts
R&S
®
NRPxxTWG
29
User Manual 1178.8371.02 ─ 04
b) Connect the R&S NRP
‑
ZK6 cable to a sensor input connector of the R&S
NRP2.
c) Connect the [RF] connector of the power sensor to the signal source.
2. Preset the R&S NRP2.
a) Press the [(PRE)SET] hardkey.
The "File" menu appears.
b) Press the [(PRE)SET] hardkey again or press the "Preset" softkey.
All parameters are set to their defaults, even when in inactive operating modes.
3. Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [ZERO] hardkey of the R&S NRP2.
The "Zero" dialog box is displayed.
c) Press the [ZERO] hardkey again to perform zeroing of all connected sensor
channels ("Zero (All)") or press the appropriate softkey to select a specific sen-
sor for zeroing.
4. Press the [FREQ] hardkey and enter the carrier frequency of the applied signal if
the specified measurement accuracy is to be reached.
5. Switch on the signal source.
The result window indicates the result (in dBm) obtained with sensor A.
6. If necessary, perform further settings.
R&S NRP2