Remote Control Commands
R&S
®
NRPxxTWG
52
User Manual 1178.8371.02 ─ 04
TRIGger:IMMediate
Causes a generic trigger event. The power sensor leaves the waiting for trigger state
immediately, irrespective of the trigger source and the trigger delay, and starts the
measurement.
This command is the only way to start a measurement if the trigger source is set to
hold (
HOLD
). Only one measurement cycle is executed, irrespective
of the averaging factor.
Usage:
Event
TRIGger:LEVel
<level>
Effective only if
.
Sets the trigger threshold for internal triggering derived from the test signal.
If you enter a value without unit, the unit is defined by
If an S-parameter device is active and/or if a mounted component with a global offset
in front of the sensor is considered, the currently effective trigger threshold and a trig-
ger threshold to be input are referenced to the appropriately shifted sensor data. If the
S-parameter device and/or the offset correction are disabled, the trigger threshold and
its input limits are adjusted as necessary.
Parameters:
<level>
Range:
300e-9 to 100.0e-3
*RST:
3.0e-5
Default unit: Watts
TRIGger:LEVel:UNIT
<unit>
Sets the unit of the trigger level if this value is entered without a unit.
See also
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
TRIGger:MASTer:PORT
<master_port>
Selects the port where the sensor outputs its own trigger event in case it is trigger mas-
ter. See
for more information.
If the sensor is the trigger master, it can output its trigger event either on the
EXTernal<1>
or
EXTernal2
.
If the sensor triggers itself, the trigger source of the sensor must be assigned to the
other external port, as shown in the examples.
Controlling the Measurement