Remote Control Commands
R&S
®
NRPxxTWG
50
User Manual 1178.8371.02 ─ 04
TRIGger:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
.
This number equals the number of results that can be obtained from the sensor after a
single measurement. As long as the defined number of measurements is not executed,
the sensor automatically initiates another measurement internally when the current
result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
See
Chapter 8.2, "Performing a Buffered Continuous Average
TRIGger:DELay
<delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
TRIGger:DELay:AUTO
<state>
If
ON
is set, no measurement is started until the power sensor
has settled. For this purpose, the delay value is automatically determined.
If a longer period is set using
, the automatically determined delay is
ignored.
Parameters:
<state>
*RST:
OFF
TRIGger:DTIMe
<dropout_time>
Sets the dropout time for the internal trigger source. During this time, the signal power
must exceed (negative trigger slope) or undercut (positive trigger slope) the level
defined by the trigger level and trigger hysteresis. At least, this time must elapse before
triggering can occur again.
See
Chapter 7.5.2.4, "Dropout Time"
Controlling the Measurement