Operating Progeny Analyzers
24
1006232_RevD
Operating Progeny Analyzers
Progeny Analyzers are durable, water-resistant, and simple to configure, providing you with the ability
to identify liquids, powders and solids in the lab, in the plant or in the field.
Before using your instrument, you may need to verify performance calibration or calibrate the
instrument. Instructions for performing those operations, as well as working with libraries, etc., are
included in this section.
Preparing to Scan: Performance Qualification
Prior to scanning materials, it is good practice to perform an Instrument Verification. Instrument
Verification is recommended if there are any issues with spectra, the instrument has been recently
serviced, or if the instrument has been stored.
The Progeny Analyzer checks for the latest calibration status to ensure that it accurately processes
spectral data. There are 2 options available for Instrumental Verification and Calibration:
1.
Performance Verification is a daily test done to ensure that the instrument is performing within
acceptable specifications. It is a check only and does not change the instrument.
2.
Instrument Calibration is done only under certain conditions as it will change instrument
parameters. These conditions include:
a.
If the calibration Status icon in the footer of the Main screen changes from
(Calibrated) to
(Not Calibrated) (Figure 10).
Figure 10. Calibrated Icon
b.
If changes have been made to the instrument.
c.
If performance of the instrument has changed.
Calibrated icon
Содержание Progeny
Страница 1: ...1006232_RevD 3 Progeny and Progeny LT Analyzer User Manual Rev D...
Страница 74: ...Advanced Options 67 1006232_RevD Figure 40 Sample Application...
Страница 106: ...Export and Import Settings 99 1006232_RevD Figure 70 Progeny Settings Screen...
Страница 111: ...Appendix B Nose Cone Positions 104 1006232_RevD Appendix B Nose Cone Positions...