Maintenance and Troubleshooting
100
1006232_RevD
9.
Maintenance and Troubleshooting
If you encounter any problems with your instrument, you’ll need to capture the Instrument Runtime
Information and Serial Number for troubleshooting.
Cleaning Your Instrument
Maintaining a clean Progeny Analyzer helps ensure that cross-substance scanning issues do not occur.
After scanning open-air samples or after the lens aperture becomes soiled, clean your instrument with
pre-packaged sanitary wipes featuring at least 70% Isopropyl alcohol cleaning solutions.
1.
Log off, remove the battery, and close the battery compartment.
2.
With 70% Isopropyl alcohol sanitary wipes, clean all loose or stuck-on materials off the surfaces
of your instrument from front to back in the following sequence:
Laser aperture
Front
Top
Sides
Bottom
Back
3.
When finished, let your instrument air dry.
Viewing Instrument Runtime Information
While powered on, Progeny Analyzers monitor runtime information for both maintaining your
instrument and complying with process regulations.
1.
Select (Information) from the Instrument Settings screen to see the color-coded status
summaries along with the following details:
Select …
To see …
Instrument
Model Number
Serial Number
Laser
Software
Version
Status
Last Calibration
Last Verification
Audit Log
Instrument status and activity data for all
users
2.
Select
(Back) to return to the Instrument Settings screen.
3.
Select
(Diagnostics) to see information related to the temperature and power:
Содержание Progeny
Страница 1: ...1006232_RevD 3 Progeny and Progeny LT Analyzer User Manual Rev D...
Страница 74: ...Advanced Options 67 1006232_RevD Figure 40 Sample Application...
Страница 106: ...Export and Import Settings 99 1006232_RevD Figure 70 Progeny Settings Screen...
Страница 111: ...Appendix B Nose Cone Positions 104 1006232_RevD Appendix B Nose Cone Positions...