Advanced Options
40
1006232_RevD
Control
Name/Type
Description
Report refresh
Refresh the current Reports view
Pan On
Tap to turn cross-hatch on
Pan off
Tap to turn cross-hatch off
Stack spectra
Stack sample and library spectra. 3 levels of stacking available
Overlay spectra
Select spectra to overlay with current spectra
Advanced Options
Custom Applications
With custom applications you can configure the instrument for specific situations, such as scanning
substances in a particular lab, facility, or location. Using a custom configuration for detection of
particular substances avoids the need to search through the Master Library.
Progeny Analyzers have two types of applications: Verification and Identification. Verification
applications check that your substances are what you expect them to be, by comparing scan results to a
specified spectral profile and then providing Pass/Fail results. Verification applications are most useful
for scanning known substances.
Identification applications identify the scanned samples by comparing to the spectral profiles in a linked
spectral sub-library, similar to Analyze, which searches the Master Library. Only users with Developer
and Administrator permission levels can add and configure applications for all users to run.
When developing applications for new substances, the user will need to optimize the instrument
settings and the focal position of the adaptor. The properties that you may need to adjust to get
sufficiently detailed spectral results are:
Содержание Progeny
Страница 1: ...1006232_RevD 3 Progeny and Progeny LT Analyzer User Manual Rev D...
Страница 74: ...Advanced Options 67 1006232_RevD Figure 40 Sample Application...
Страница 106: ...Export and Import Settings 99 1006232_RevD Figure 70 Progeny Settings Screen...
Страница 111: ...Appendix B Nose Cone Positions 104 1006232_RevD Appendix B Nose Cone Positions...