Operating Progeny Analyzers
34
1006232_RevD
Table 1. Available Properties for Analyze and Application Scans
Categories/Properties
Control
Possible Values
Description
Collect Settings
Auto Collect
Soft Key
Auto Power Setting
from 30-490 mW
Note: Progeny LT
range is 30-300 mW
Power settings is determined automatically by
analyzing the spectral response of the sample
Note:
When Auto Collect is enabled (ON) Power
settings is disabled.
Power
Soft Key
300 mW
(default)
Range:
30 mW (min)
490 mW (max)
300 mW for LT
Opens a settings screen through which you can
scroll to select the value.
Settings available at increments of 5 mW.
Auto Exposure
Soft Key
Auto Exposure
range from 50-
30,000 ms
Exposure time and Averages is determined
automatically by analyzing the spectral response of
the sample
Note:
When Auto Exposure is enabled (ON), Exposure
and Averages settings are disabled.
Exposure
Soft Key
1,000 ms
(default)
Range:
5 ms (min)
30,000 ms
(max)
Opens a settings screen through which you can
scroll to select the value.
Settings available at the following increments:
10 ms increments from 10 — 1,000 ms
100 ms increments from 1,000 — 30,000
ms
Averages
Soft Key
1 count
(default)
Range:
1 count (min)
30 count (max)
Opens a settings screen through which you can
scroll to select the value.
Settings available at Increments of 1
count.
Take Background
Every Scan
Switch
(On—
default)
(Off)
This feature minimizes the effect of possible
instrument drift on scan results.
Default is
(On).
When On, the instrument will collect a
new dark background spectrum with
every measurement and subtract that
from the spectrum collected with the
laser on.
When
(Off), the instrument will
use the most recently collected dark
background spectrum.
Содержание Progeny
Страница 1: ...1006232_RevD 3 Progeny and Progeny LT Analyzer User Manual Rev D...
Страница 74: ...Advanced Options 67 1006232_RevD Figure 40 Sample Application...
Страница 106: ...Export and Import Settings 99 1006232_RevD Figure 70 Progeny Settings Screen...
Страница 111: ...Appendix B Nose Cone Positions 104 1006232_RevD Appendix B Nose Cone Positions...