Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-76 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
Place the beam alignment tool to the sensor alignment tool to vertical position as shown in the
figure below.
Figure 142
Stand behind the tube head and protect yourself from radiation. Press and hold down the
exposure button. The image of the radiation beam will appear on the beam alignment tool.
The beam image should appear within the borders of the rectangle marked on the beam
alignment tool. If it does not adjust the second primary collimator position. Refer to section 7.8
“Second primary collimator adjustment” on page E-78.
Figure 143
Move the beam alignment tool to the upper position on the sensor alignment tool as shown in
the figure above.
Beam alignment tool
D2filt_D
.eps
Beam alignment
Beam alignment
tool in upper
position
tool in lower
position
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......