Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-17
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Adjust the head support angle with the two screws located on the left side of the mechanism
(use 2.5mm allen key) (see Fig. 29 on page E-16 and Fig. 30 below).
Figure 30
Attach the removed covers.
Primary slot adjustment - radiation beam too low or too high
WARNING
DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE X-RAYS
ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION WITH
THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM IS
STILL MISALIGNED REPEAT THE OPERATION.
If the beam is too high or low loosen the two screws that hold the outer slot in position and
adjust as necessary. Tighten the screws and recheck the beam. Repeat the above procedure
until the beam is correctly positioned.
Figure 31
Spirit level
90°
Beam too low
Securing screws
Beam too high
Outer collimator
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......