Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-6 Planmeca Proline XC X-ray unit
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Loosen the lock disc attachment screw with the 3mm allen key (Fig. 11, 1) and remove the
lock disc from the adapter (Fig. 11, 2).
Figure 11
Push the button axle inwards and remove the sensor head from its adapter.
Figure 12
The sensor alignment tool can now be attached to the adapter pins.
1
2
Sensor head
Button axle
Содержание Proline XC