Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-58 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
Cephalostat sensor head with quick connector mechanism -
removing the covers
Unscrew the bottom cover plate attachment screw with the 4mm allen key and remove the
cover plate.
Figure 110
Turn the lock disc from above the locking plate and push in the quick connector mechanism
locking knob. Unscrew the attachment screw of the quick connector mechanism cover with
the 4mm allen key.
Figure 111
Bottom cover plate
Cover
Locking plate
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......