Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-89
CEPHALOSTAT ADJUSTMENT
Technical Manual
The system suggests default kV and mA values. The default kV value for the beam check is
70 and the default mA value is 10. Touch the
Ready
field.
Figure 171
NOTE
In case the contrast on the exposure is not satisfactory, change the values by touching
the kV/mA field and take a test exposure.
The following window appears on the computer screen.
Figure 172
Press the X-ray unit’s exposure switch.
After the exposure is taken the following window appears.
Figure 173