Chapter D - ADJUSTMENT & CALIBRATION
D-46 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.3 Head support adjustment
Switch off the unit. Remove the cephalostat arm upper cover, attachment casting cover and
attachment mechanism cover.
Make sure that the locking screw of the adjustment flange is loosened (Fig. 76, 1). Adjust the
head support position according to the Fig. 76. Rotate the head support around its vertical
axis to adjust the ear posts in horizontal direction (Fig. 76, 2) and around its horizontal axis to
adjust the ear posts in vertical direction (Fig. 76, 3).
If the cephalostat cassette holder does not run parallel with the primary collimator, adjust the
cassette holder in vertical direction (Fig. 76, 4).
Figure 76
Fasten the loosened screws and replace the covers.
Remove the head support alignment tools and place the ear posts in the ear post holders.
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
1
Soft tissue filter adjustment
(see section “Soft tissue filter adjustmen
2
3
4
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......