Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-11
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Make sure that the X-ray unit is in the panoramic exposure mode (text Pan on the main dis-
play). Touch the ready field to drive the X-ray unit into the exposure position. The rotating unit
will move to ready position.
Figure 21
On the exposure switch holder a ready indicator light will come on and on the main display the
Ready field will be green. The Dimaxis program will show the
Waiting for Ready
message on
the computer screen.
The indicator light will flash while the unit moves to the ready position. They will stop flashing
when the unit has reached the ready position for the selected panoramic exposure program.
The Dimaxis program will now show the
Waiting for Exposure
message on the computer
screen.
NOTE
The ready indicator light will be flashing if the PC is not in the ready state. Refer to the
Dimaxis user’s manual on how to prepare the system ready for the exposure.
Select the Test mode as follows. Touch the i-field on the
Main
display. The list of User set-
tings appears.
Figure 22
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......