Chapter D - ADJUSTMENT & CALIBRATION
D-20 Planmeca Proline XC X-ray unit
PATIENT POSITIONING MECHANISM
Technical Manual
If the distance from the middle ball to the center of the tenth ball on the left-hand side is less
than 118mm and on the right-hand side more than 122mm then the patient positioning mech-
anism must be moved towards the column.
Figure 37
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards.
Figure 38
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be
moved backwards.
Figure 39
Right-hand side
Film
Less than 118mm
More than 122mm
Film
More than 122mm
More than 122mm
Right-hand side
Less than 118mm
Less than 118mm
Right-hand side
Film
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......