Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-90 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
You can type any kind of additional information (e.g. name of the clinic) on this window. Con-
firm by clicking the
OK
button.
The test report appears automatically and is shown as html page.
Figure 174
NOTE
The report can be printed or saved in the computer.
NOTE
Check that the radiation beam can be seen entirely on the image receptor. The radia-
tion beam should not exceed the receptor.
If the radiation is not shown on the image receptor, calibrate the X-ray unit again and take
another exposure.
If the test report shows that the radiation can be seen on the image receptor, you can exit the
Proline Beam Check program by selecting
Exit
from the
File
pulldown menu.
Test R eport, Thu Jun 10 09:07:21 2004
Date and time according
to the PC
Cephalometric Test Report
Detailed view of the upper part
of the active area
Detailed view of the lower part
of the active area
Beam location in the active
area of the sensor
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......