
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-87
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.10 Beam check
The beam check feature allows you to check that the X-ray unit is properly collimated. We
recommend that you perform the beam check after the normal cephalostat beam adjustment.
To run the Beam Check test, proceed as follows:
Start the Beam Check program by clicking
Start
>
Programs
>
Planmeca.
Select
Beam Check
from the appearing list.
Figure 166
The window shown below appears on your computer. Select Proline Ceph from the X-ray
equipment list. Click
Take the Image
button.
Figure 167
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......