Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-40 Planmeca Proline XC X-ray unit
PATIENT POSITIONING MECHANISM
Technical Manual
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards (Fig. 82).
Figure 82
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be
moved backwards (Fig. 83).
Figure 83
ST
AR
T
EN
D
PL
AN
ME
CA
No
.5
09
71
X-
LIN
E
Y-
LIN
E
P
a
tfron.eps
More than 122 mm More than 122 mm
Right-hand side
Right-hand side
ST
AR
T
EN
D
PL
AN
ME
CA
No
.5
09
71
X-
LIN
E
Y-LINE
P
atpac
k.eps
Less than 118 mm Less than 118 mm
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......