Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-22 Planmeca Proline XC X-ray unit
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
If you cannot adjust the beam far enough to the left using the method described above you
may have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab
slightly and then tighten the screw. The collimator position should then be rechecked.
Figure 40
Touch the Next field to store the collimator position in the memory and move on to the next
collimator slot. The number of the next slot will appear on the display.
Repeat the alignment procedure described above for the other collimator slots (1, C, d).
Exit the primary collimator calibration mode. Exit the service settings mode.
Metal tab
Positioning thread
Draw nut