
Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-15
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
If you cannot adjust the beam far enough to the left using the method described above you
may have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab
slightly and then tighten the screw. The collimator position should then be rechecked.
Figure 26
Touch the Next field to store the collimator position in the memory and move on to the colli-
mator number
1
. The number 1 will appear on the display.
Figure 27
Repeat the alignment procedure described above for the collimator number 1.
Exit the primary collimator calibration mode. Exit the service settings mode.
Metal tab
Positioning thread
Draw nut
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......