Chapter D - ADJUSTMENT & CALIBRATION
D-2 Planmeca Proline XC X-ray unit
REQUIRED TOOLS
Technical Manual
•
Ball phantom (part number 50971). For checking the position of the patient positioning
mechanism and the positioning lights.
Figure 2
•
Frankfort plane alignment tool (part number 50977). Used with the ball phantom for check-
ing the position of the Frankfort plane light.
Figure 3
•
Beam alignment tool (part number 50972). For checking the position of the panoramic X-
ray beam.
Figure 4
START
END
PLANMECA
No.50971
X-LINE
Y-LIN
E
Ball phantom
START
END
PLANM
ECA
No.50971
X-LINE
Y-LINE
PLANMECA OY
NO.50977
Frankfort plane
alignment tool