Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-17
PATIENT POSITIONING MECHANISM
Technical Manual
Slide the cassette into the cassette holder. Make sure that you push it all the way in.
The exposure values 60 kV and 2 mA must be selected to produce a clear ball phantom expo-
sure and it is only possible to select these values in the Test exposure mode. Enter the serv-
ice settings mode. From the list of service settings select Test exposure.
Figure 31
The display shown below appears. Touch the
kV
/
mA
field.
Figure 32