Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-41
PATIENT POSITIONING MECHANISM
Technical Manual
4.2
Patient positioning mechanism adjustment
Adjustment in ball phantom’s y-direction
If the patient positioning mechanism is too near or too far from the column it will have to be
centered.
Remove the cover from the lower shelf. Replace the ball phantom tool.
Remove the sensor head from the quick connector mechanism (see instructions given in
Planmeca Proline XC panoramic X-ray unit’s user’s manual). Attach the sensor alignment tool
to the quick connector.
Select the Test mode as follows. Touch the i-field on the
Main
display. The list of User set-
tings appears.
Figure 84
From the list of User settings select
Behavioural preferences
. The display shown below
appears. Touch the Test mode field to enable the test mode (ON).
Figure 85