Chapter G - TROUBLESHOOTING
G-12 Planmeca Proline XC X-ray unit
DISTURBANCES IN TEMPLE SUPPORT MECHANISM
Technical Manual
8 DISTURBANCES IN TEMPLE SUPPORT MECHANISM
8.1 Temple supports do not move
8.2 Temple supports function erroneously
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Motor does not start
Keyboard out of order.
Motor out of order or connector
loose.
PCB out of order or connector
loose.
Motor starts but supports
immobile
Pin loose.
Drawing nut glides on motor
axle.
Thread of drawing nut dam-
aged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Temple supports do not close
properly
Attachment of quadrants to
temple support cranks has
yielded.
Adjustment of magnet and Hall
sensor erroneous.
Temple supports do not move
symmetrically
Attachment of quadrants to
temple support cranks has
yielded.
Temple supports have too
much clearance.
Adjustment of clearance incor-
rectly done.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......