Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-81
CEPHALOSTAT ADJUSTMENT
Technical Manual
The calibration must be performed with 3x3 (enhanced resolution) and 4x4 (normal resolu-
tion) binning. With enhanced resolution select
60
kV and
11
mA, and with normal resolution
select
60
kV and
6
mA. Note that these values must be selected to produce a clear exposure.
NOTE
The 2 mA value can be selected in the test exposure mode, refer to section 2.2 “Test
exposure” on page B-10.
Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon
located in the Dimaxis program folder. The Dimax2Tool window appears.
Figure 150
Calibrating the sensor head
Select Proline from the Settings / Type pulldown menu.
Figure 151
The calibration must be performed with both 3x3 (enhanced resolution) and 4x4 (normal res-
olution) binning. Select the binning from the Settings / Binning pulldown menu.
Figure 152
Click the Dimax3 cephalometric exposure button. The pmpcal2 window with text Waiting
appears.
Figure 153
Dimax3 cephalometric
exposure
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......