Chapter A - GENERAL & TECHNICAL DATA
Planmeca Proline XC X-ray unit A-15
EMC INFORMATION
Technical Manual
Recommended separation distances between
portable and mobile RF communications equipment and Planmeca Proline XC X-ray unit
Planmeca Proline XC X-ray unit is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of Planmeca Proline XC X-ray unit can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Planmeca Proline XC X-ray unit as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01
0.2
0.2
0.3
0.1
0.4
0.4
0.7
1
1.2
1.2
2.4
10
4.0
4.0
8.0
100
12.0
12.0
24.0
d
1,2 P
=
d
1,2 P
=
d
2,3 P
=
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......