Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-37
AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRATION
Technical Manual
Move the cassette carriage to the loading position by touching the Entry field.
Figure 63
In the darkroom open the cassette and place a film inside the cassette. Close the cassette.
Slide the cassette into the cassette holder.
NOTE
The film density adjustment must be performed in normal panoramic mode. Make sure
that the unit is NOT in the AEC mode.
Select a kilovolt value of 68 and a milliampere value of 7.
Drive the X-ray unit to the exposure position by touching the Ready field. The rotating unit will
turn to the start position and the indicator light will come on.
Figure 64
Take an exposure.
Touch the Entry field to drive the cassette to the loading position, remove the cassette and
process the film in the same way you would do in clinical exposure procedure.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......