Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-13
PANORAMIC X-RAY BEAM ADJUSTMENT
Technical Manual
Radiation beam too far to the right or left
WARNING
WARNING: DO NOT ADJUST THE POSITION OF THE PRIMARY COLLIMATOR MECHA-
NISM WHILE X-RAYS ARE BEING GENERATED. CHECK THE BEAM POSITION,
ADJUST THE POSITION WITH THE X-RAYS SWITCHED OFF AND THEN RECHECK THE
POSITION. IF THE BEAM IS STILL MISALIGNED REPEAT THE OPERATION.
NOTE
THE CEPH PRIMARY COLLIMATORS ARE ALSO ADJUSTED ACCORDING TO THE
INSTRUCTIONS GIVEN IN THIS SECTION.
If the beam is too far to the left or right of the alignment rectangle it must be centered.
Figure 22
NOTE
A microprocessor controlled motor determines the position of each collimator. If the
beam is readjusted to the left or right the new position must be reprogrammed into the
memory.
Enter the service settings mode if you are not already in it.
From the list of Service settings select Primary collimator.
Figure 23
Beam too far to the right
Beam too far to the left
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......