Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-78 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.8
Second primary collimator adjustment
Second primary collimator in too low or high position
Loosen all the three screws located on the side of second primary collimator with the 4mm
allen key.
Figure 144
Move the second primary collimator in vertical direction until the collimator is correctly posi-
tioned. Tighten the loosened screws.
Second primary collimator not vertical
Loosen the two screws located on the side of second primary collimator.
Figure 145
Loosen these three screws.
Loosen these two screws.