10 Maintenance
C-863.12 Mercury Controller
MS249E
Version: 1.2.1
251
−
Select the
C-863.12
entry in the
Select Controller:
field.
−
Select the COM port of the PC connected to the C-863.12 in the
Select COM port:
field.
−
Do not change the
Select Firmware:
field. The "DSP" (DSP = Digital Signal Processor)
is automatically entered by selecting the controller.
5.
Select the new firmware file:
a)
Click the
Select
button.
b)
Go to the directory in the file selection window where you stored the firmware file.
c)
Double-click the new firmware file (.hex extension) to enter the file path in the
Filename
field.
6.
Start the firmware update by clicking on the
Start Update
button.
The firmware of the C-863.12 is updated. The update progress is displayed in the
message list and by the progress bar.
The update was successful when the
disconnect Comport
message appears as
the last entry in the message list.